Willowa 40 Street, 05-205 Dobczyn · gazmed@gazmed.pl
Facebook
LinkedIn
YouTube

OXYKON-DUO oxygen concentrator

OXYKON-DUO – Dependable and economic alternative for traditional supply sources of oxygen in medical facilities.

OXYKON oxygen concentrator

OXYKON-DUO oxygen concentrator - basic information

Dependable and economic alternative for traditional supply sources of oxygen in medical facilities.

OXYKON-DUO Concentrator based supply system can be easily to adapted to existing system of supply and allows to enjoy significantly reduced costs of oxygen therapy.

 

Production of oxygen is based on adsorption properties of molecular sieves. Compressed air produced by air compressors is distributed to production modules, where nitrogen and other constituents of air are removed by molecular sieve, and oxygen as a final product is being stored in buffer tank, concentration of produced gas is even 95% +/- 1% of oxygen. Rest is mainly argon and other inert gases naturally existing in air.

Researches and our experience confirm that such concentration covers 99% of hospital needs regarding oxygen therapy. Especially when it applies to mechanical ventilation of patients treated in ICU wards, pulmonology diseases, during surgical procedures and many others where oxygen is necessary for patient therapy.

Safe and dependable system

Oxygen concentrator OXYKON is a device built and working according to PSA method of operation. It is made to work continuously producing unlimited amounts of oxygen. Depending on amounts of connected modules it covers the demand in small and big clinics. It is also used in mobile container systems used for example in field military hospitals.

Control over system is possible through control panel, which depending to demand switches on and off production modules, also controls work of the cylinder reserve system, and when necessary switches it on to provide continuous supply of oxygen. Oxygen concentration is monitored continuously, to provide product at the highest quality. As an option it is also possible to control ambient concentration in the room where OXYKON-DUO is installed in. Any abnormal situations are controlled by our system and being transferred to alarm unit installed in control room and by SMS to persons responsible for maintaining the system.

System OXYKON comprises to requirements of:

  • EN ISO 7396-1:2016.07 describing the requirements for oxygen concentrator systems used with central piping system in hospitals regarding design, installation, construction, work monitoring and maintaining this systems.
  • EN ISO 13485:2016 Quality assurance system for medical device.
  • MDD 93/42/EU
  • and many others.

Economical system, guarantying independence and optimal usage of space.

Oxygen concentrator OXYKON allows medical facilities to be independent from outside gas deliveries and changing prices which allows to reduce oxygen therapy costs by up to 80%. To produce 1 Nm3 of oxygen system requires only 0,8 – 1,2 kW electrical power. Production costs are stable and very easy to calculate. Service costs are very low and limited mainly to compressors delivering compressed air. Hospitals already using our system had lowered costs significantly and become fully independent from whims of gas delivering firms.

Single oxygen concentrator may have capacity from 40 up to 1200 liters per minute with working pressure up to 6 bar. This corresponds to a demand of 9 to 270 cylinders with a capacity of 40 liters per day with a filling pressure of 150 bar or from 78 to 2350 kg of liquid oxygen per day.

Cylinder filling system

As an additional option system can fill cylinders up to the pressure of 150-200 bars.

From medical point of view our oxygen concentrator system is able to cover all needs regarding oxygen therapy in any hospital. Mixture of gases delivered to patient is always a mixture of oxygen and air in different proportions, whether the patient breathes by himself or through machine.

There is no medical arguments for using oxygen 99,9% to prepare a mixture that is usually in concentration of 30-60%. It is also wrong and unhealthy from medical point of view presenting to the patients 100% of oxygen, or using very high concentrations of oxygen, it can cause irreversible damage to the lungs tissue.

Produced gas control

Concentrator is equipped in system of continuous control with log, also it is equipped in safety feature that will switch it off and switch to reserve supply if produced gas doesn’t meat required parameters or when the pressure will drop below safety threshold. Produced gas after it passes through molecular sieve has dew point of -70ºC. Maintenance is limited only to air compressors that is conducted by dedicated compressor producer service. Oxygen concentrator modules require almost no maintenance.

If you decide to choose our solution, Hydro-Gaz-Med will lead the whole process from design, through installation to commissioning of the system.

capacity: from 40 to 1200 liters of gasous oxygen per minute per OXYKON-DUO unit

or from 3.2 up to 98 kg liquid oxygen per hour

Model No. Outlet pressure [barg] Capacity m3/h Air consumption Air/O2
93% 95% +/-1% m³/min m³/h
DUO40 4 2,44 2,27 0,41 25 10,16
5 2,58 2,39 0,47 28 11,01
6 2,64 2,45 0,50 30 11,28
DUO60 4 3,49 3,23 0,58 35 9,97
5 3,68 3,41 0,66 40 10,81
6 3,78 3,50 0,70 42 11,07
DUO70 4 5,32 4,43 1,02 61 11,56
5 5,61 4,73 1,17 70 12,53
6 5,76 4,85 1,23 74 12,83
DUO100 4 5,91 5,32 1,13 68 11,50
5 6,80 6,21 1,36 82 12,00
6 7,45 6,80 1,53 92 12,30
DUO190 4 12,41 11,23 2,07 124 10,00
5 13,00 11,52 2,27 136 10,45
6 15,37 14,18 2,96 177 11,54
DUO280 4 19,50 16,84 3,35 201 10,30
5 22,16 20,09 4,04 242 10,93
6 24,48 22,46 4,51 271 11,06
DUO430 4 23,72 21,09 4,25 255 10,75
5 28,27 25,97 5,17 310 10,98
6 30,05 26,90 5,78 347 11,54
DUO640 4 35,46 31,52 6,35 381 10,75
5 42,26 38,81 7,73 464 10,98
6 44,92 40,21 8,64 518 11,54
DUO880 4 46,79 41,86 7,83 470 10,04
5 51,52 46,20 9,01 541 10,50
6 56,15 50,24 10,64 638 11,37
DUO1250 4 64,03 56,15 10,64 638 9,97
5 70,43 61,76 12,31 739 10,49
6 76,83 67,37 14,38 863 11,23
DUO1500 4 73,88 64,03 12,31 739 10,00
5 81,26 70,43 14,18 851 10,47
6 88,65 76,83 16,65 999 11,27

Standard equipment covers:

  • OXYKON-DUO unit;
  • buffer tank;
  • set of pressure reducers;
  • dew point and carbon monoxide monitor.

Optional equipment:

  • compresssed air system with conditioning system;
  • oil indication system;
  • oxygen monitor with paramagnetic oxygen cell, electronic flow monitor and pressure.